Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 1 de 1
Filtrar
Mais filtros










Base de dados
Intervalo de ano de publicação
1.
Health Phys ; 99(4): 453-6, 2010 Oct.
Artigo em Inglês | MEDLINE | ID: mdl-20838085

RESUMO

During elution of 99Mo-99mTc generators used in nuclear medicine, 99Mo might be extracted becoming a radionuclidic impurity. According to the International Atomic Energy Agency, the activity ratio between 99Mo and 99mTc in the eluate, at the moment of administration to the patient, should not exceed 0.015%. The aim of this work is to optimize a methodology to determine 99Mo activity in 99mTc eluates. Efficiency curves were obtained using a NaI(Tl)8"×4" scintillation detector. The methodology was validated by measuring a standard solution of 99Mo. It was concluded that the technique is sensitive to detect 99Mo in 99mTc eluates at levels below international limits.


Assuntos
Molibdênio/análise , Geradores de Radionuclídeos , Cintilografia/instrumentação , Contagem de Cintilação/instrumentação , Pertecnetato Tc 99m de Sódio/análise , Calibragem , Limite de Detecção , Cintilografia/métodos , Reprodutibilidade dos Testes , Contagem de Cintilação/métodos , Pertecnetato Tc 99m de Sódio/química
SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA
...